Medtronic N'Vision Clinician Programmer 8840 N'Vision Clinician Programmer and 8870 N'Vision removable application card does not encrypt PII and PHI while at rest.
Configurations

Configuration 1 (hide)

AND
cpe:2.3:o:medtronic:n\'vision_8840_firmware:-:*:*:*:*:*:*:*
cpe:2.3:h:medtronic:n\'vision_8840:-:*:*:*:*:*:*:*

Configuration 2 (hide)

AND
cpe:2.3:o:medtronic:n\'vision_8870_firmware:-:*:*:*:*:*:*:*
cpe:2.3:h:medtronic:n\'vision_8870:-:*:*:*:*:*:*:*

History

22 May 2025, 18:15

Type Values Removed Values Added
CWE CWE-311
References
  • () http://www.securityfocus.com/bid/104213 -
  • () https://global.medtronic.com/xg-en/product-security/security-bulletins/nvision.html -
Summary (en) Medtronic N'Vision Clinician Programmer 8840 N'Vision Clinician Programmer, all versions, and 8870 N'Vision removable Application Card, all versions. The 8840 Clinician Programmer executes the application program from the 8870 Application Card. An attacker with physical access to an 8870 Application Card and sufficient technical capability can modify the contents of this card, including the binary executables. If modified to bypass protection mechanisms, this malicious code will be run when the card is inserted into an 8840 Clinician Programmer. (en) Medtronic N'Vision Clinician Programmer 8840 N'Vision Clinician Programmer and 8870 N'Vision removable application card does not encrypt PII and PHI while at rest.
CVSS v2 : 4.6
v3 : 6.6
v2 : 4.6
v3 : 4.6

21 Nov 2024, 03:41

Type Values Removed Values Added
References () https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01 - Third Party Advisory, US Government Resource () https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01 - Third Party Advisory, US Government Resource
References () https://www.medtronic.com/security - Vendor Advisory () https://www.medtronic.com/security - Vendor Advisory

19 Aug 2024, 19:35

Type Values Removed Values Added
CWE CWE-284
CVSS v2 : 4.6
v3 : 6.8
v2 : 4.6
v3 : 6.6

Information

Published : 2018-07-13 19:29

Updated : 2025-05-22 18:15


NVD link : CVE-2018-10631

Mitre link : CVE-2018-10631

CVE.ORG link : CVE-2018-10631


JSON object : View

Products Affected

medtronic

  • n\'vision_8840_firmware
  • n\'vision_8870
  • n\'vision_8870_firmware
  • n\'vision_8840
CWE
CWE-311

Missing Encryption of Sensitive Data

CWE-693

Protection Mechanism Failure

CWE-284

Improper Access Control