CVE-2018-8868

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.
Configurations

Configuration 1 (hide)

AND
cpe:2.3:o:medtronic:24950_mycarelink_monitor_firmware:-:*:*:*:*:*:*:*
cpe:2.3:h:medtronic:24950_mycarelink_monitor:-:*:*:*:*:*:*:*

Configuration 2 (hide)

AND
cpe:2.3:o:medtronic:24952_mycarelink_monitor_firmware:-:*:*:*:*:*:*:*
cpe:2.3:h:medtronic:24952_mycarelink_monitor:-:*:*:*:*:*:*:*

History

22 May 2025, 19:15

Type Values Removed Values Added
CVSS v2 : 6.9
v3 : 6.4
v2 : 6.9
v3 : 6.2
References
  • () https://global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-6-28-18.html -
Summary (en) Medtronic MyCareLink Patient Monitor, 24950 MyCareLink Monitor, all versions, and 24952 MyCareLink Monitor, all versions, contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can apply the other vulnerabilities within this advisory to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality. (en) Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.

21 Nov 2024, 04:14

Type Values Removed Values Added
References () https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01 - Third Party Advisory, US Government Resource () https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01 - Third Party Advisory, US Government Resource

Information

Published : 2018-07-03 01:29

Updated : 2025-05-22 19:15


NVD link : CVE-2018-8868

Mitre link : CVE-2018-8868

CVE.ORG link : CVE-2018-8868


JSON object : View

Products Affected

medtronic

  • 24952_mycarelink_monitor_firmware
  • 24952_mycarelink_monitor
  • 24950_mycarelink_monitor_firmware
  • 24950_mycarelink_monitor
CWE
CWE-749

Exposed Dangerous Method or Function

NVD-CWE-noinfo