Vulnerabilities (CVE)

Filtered by vendor Medtronic Subscribe
Filtered by product 2090 Carelink Programmer
Total 3 CVE
CVE Vendors Products Updated CVSS v2 CVSS v3
CVE-2018-5448 1 Medtronic 2 2090 Carelink Programmer, 2090 Carelink Programmer Firmware 2025-05-22 2.7 LOW 4.8 MEDIUM
Medtronic 2090 CareLink Programmer’s software deployment network contains a directory traversal vulnerability that could allow an attacker to read files on the system.
CVE-2018-5446 1 Medtronic 2 2090 Carelink Programmer, 2090 Carelink Programmer Firmware 2025-05-22 2.1 LOW 4.9 MEDIUM
Medtronic 2090 CareLink Programmer uses a per-product username and password that is stored in a recoverable format.
CVE-2018-10596 1 Medtronic 2 2090 Carelink Programmer, 2090 Carelink Programmer Firmware 2025-05-22 5.2 MEDIUM 7.1 HIGH
Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.